The device history records, including records of acceptance testing, were reviewed for both devices and found to conform to all process and device specifications.The returned devices were visually inspected in detail and dimensional measurements were made.Both devices had been subject to extensive repair and modification by unauthorized third parties and showed evident signs of extreme wear and tear.Material had been forcefully removed from several parts of the instruments by grinding, reducing the original dimensions of the devices and compromising their cutting function as well as their load capacity.In addition to the above, the individual devices had been specifically damaged as follows: device #1, lot# 27156613: the shape of the device was changed by removal of an essential part required to ensure mechanical safety and functional effectiveness of moving parts of device.The movement of the sliding parts of the device was thus extended and caused additional damage to the tip.Device #2, lot# 27161331: the cutting edge was bent upwards.The device is intended for cutting of bony tissue.However, the tip showed damage typically caused by grasping hard objects such as wire.
|
During a surgical case, two kerrison rongeurs malfunctioned and were believed to have fractured.By means of an x-ray, it was confirmed that no part of the instruments had fallen into the patient's body.The patient suffered no impact and the procedure was completed as planned.
|