• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 5392
Device Problems Device Displays Incorrect Message (2591); Operating System Becomes Nonfunctional (2996)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/15/2016
Event Type  Injury  
Manufacturer Narrative
This device was reported as included in the field action.Based on the information received and without the return of the product, it cannot be confirmed that this device performed as described in the field action.It is included in the field action in the abundance of caution.
 
Event Description
It was reported that the external pulse generator (epg) was in use with a patient when it stopped functioning and displayed an error.The epg was replaced and is expected to be returned.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: analysis confirmed the reported issue; the device powers up to an error and the main printed circuit board (pcb) is electrically out of specification.Additional errors were found in the logs.It was also found that one case screw was contaminated.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the external pulse generator (epg) was in use with a patient when it stopped functioning and displayed an error.The epg was replaced and has been returned for repair.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Failure analysis was performed on the main board.Visual inspection revealed no anomalies.Benchtop analysis found that the unit powered on to an error.After the eeprom (electrically erasable programmable read-only memory) was replaced, all tests passed on the automated test console.Errors were also displayed in the logs.Conclusion: the reported event was confirmed, the failure was caused by a corrupt eeprom.One of the errors was confirmed in the log but could not be confirmed on the bench.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6320510
MDR Text Key67080705
Report Number3004593495-2017-00087
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K150246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 03/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number5392
Device Catalogue Number5392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/04/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/03/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-1356-2015
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
Patient Weight60
-
-