Model Number S7
Device Problem No Display/Image (1183)
Patient Problem No Patient Involvement (2645)
Event Date 09/08/2016
Event Type  malfunction  
Manufacturer Narrative
The initial report occurred on (b)(6) 2016 and was found to be a non-reportable malfunction based on the information available at the time the event occurred.On (b)(6) 2017, a retrospective review related to a capa was completed for all events where the reported malfunction was not previously associated with serious injuries; the reporting determination has been revised to consider these malfunction events reportable.No patient information provided as no patient was involved in this concern.Device manufacturing date is unavailable.Onsite investigation was able to replicate the reported event.Replacement of the surgeon monitor along with the monitor cable resolved the issue.No further issues were reported.Analysis of the returned cable could not confirm any failure as it passed bench testing and functioned as intended and without issues.Analysis of the returned monitor confirmed an electrical failure as the issue was reproduced during testing.A full system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A site representative reported that when they plugged in the navigation system's surgeon monitor it stayed on for a couple of minutes then shut off.There was no patient present when this issue was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
sam saleh
826 coal creek circle
louisville, CO 80027
MDR Report Key6321919
MDR Text Key67331897
Report Number1723170-2017-00636
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/16/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse