Model Number R SERIES |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device's display showed multiple colored lines and no clinical data could be seen.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
The device was not returned to zoll medical corporation; the device's lcd module was removed and returned.The malfunction was duplicated and attributed to a faulty lcd module.Analysis for reports of this type has not identified an increase in trend.
|
|