A portion of the lead assembly was not returned for analysis; therefore a complete evaluation could not be performed on the entire lead product.During the visual analysis the connector pin quadfilar coil appeared to be broken approximately 185 mm and 192 mm from the end of the connector boot.Scanning electron microscopy was performed on the connector pin quadfilar coil break and the area was identified as having evidence of a stress induced fracture (fatigue appearance) with mechanical damage, fine pitting on one of the broken coil strands and evidence of a stress induced fracture (rotational forces) which most likely completed the fracture on two of the broken coil strands.Scanning electron microscopy was performed on the connector pin quadfilar coil break and identified the area as being mechanically damaged with pitting which prevented identification of the coil fracture type.Pitting was observed on the coil surface.It is believed that stimulation was present for a certain period of time as evidenced by the presence of metal pitting.Low magnification sem analysis of the quadfilar coil shows characteristics typical of a lead discontinuity which may include: material fracture, rough or pitted surface, thinned material thickness, electro-etching or material dissolution.The abraded openings found on the outer silicone tubing and the cut ends that were made during the explanted process, most likely provided the leakage path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.With the exception of the observed discontinuity the condition of the returned lead portion is consistent with conditions that typically exist following an explant procedure.No other obvious anomalies were noted.The set screw marks found on the lead connector pin provide evidence that, at one point in time, a good mechanical and electrical connection was present.Continuity checks of the returned lead portion were performed, during the visual analysis, and no other discontinuities were identified.Based on the findings in the lab, there is evidence to suggest a discontinuity in the returned portion of the device which may have contributed to the stated high impedance.The pulse generator was explanted/returned due to ¿prophylactic replacement¿.Various electrical loads were attached to the pulse generator and results of diagnostic tests demonstrate that accurate resistance measurements were obtained in all instances.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.
|