• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HERAEUS KULZER GMBH GLUMA DESENSITIZER; AGENT, TOOTH BONDING, RESIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HERAEUS KULZER GMBH GLUMA DESENSITIZER; AGENT, TOOTH BONDING, RESIN Back to Search Results
Catalog Number 65872354
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Tissue Damage (2104); Caustic/Chemical Burns (2549)
Event Date 01/23/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).This is a serious injury (as defined in 21 cfr section 803.3) as the patient reported having an adverse reaction.Because the malfunction allegation could not be confirmed, the cause of the adverse reaction could not be determined.This incident is being reported to maintain compliance with 21 cfr 803 and out of an abundance of caution.Directions for use indicate rubber dam use is required.Isolation of the patient's soft tissue was not maintained.No product returned.
 
Event Description
(b)(6) called to report a patient complaining about her entire mouth peeling and sore.(b)(6) had placed gluma on entire mouth for one minute the day before, not maintaining isolation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLUMA DESENSITIZER
Type of Device
AGENT, TOOTH BONDING, RESIN
Manufacturer (Section D)
HERAEUS KULZER GMBH
philipp-reis-strasse 8/13
wehrheim, D-612 73
GM  D-61273
Manufacturer (Section G)
HERAEUS KULZER GMBH
philipp-reis-strasse 8/13
wehrheim, D-612 73
GM   D-61273
Manufacturer Contact
rita rogers
300 heraeus way
south bend, IN 46614
5742995409
MDR Report Key6333102
MDR Text Key67534120
Report Number9610902-2017-00002
Device Sequence Number1
Product Code KLE
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K962812
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Dental Hygienist
Type of Report Initial
Report Date 01/24/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Dental Hygienist
Device Catalogue Number65872354
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date01/24/2017
Date Manufacturer Received01/24/2017
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age35
-
-