• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MEXICO PACIFIC XTREME; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MEXICO PACIFIC XTREME; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Catalog Number PCF030150130
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Intimal Dissection (1333)
Event Date 01/18/2017
Event Type  Injury  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Physician was attempting to treat a lesion in the peroneal artery.A 3mmx150mm balloon size was requested.The box was opened prior to use and the pouch labelling was not checked against the carton of the device.The balloon was prepped and inserted as per ifu.During inflation the surgeon realized the balloon appeared bigger than the requested 3mm balloon.When reviewing the packaging and the inner pouch it was then realized that the device was labelled as a 5mmx150mm.The balloon was removed and another 3mmx150mm was used.A xience stent was required to treat a dissection of the peroneal artery.The surgeon was unsure if this was caused by the 5mm x150mm balloon inflation.Post procedure good run off to distal vessels.Mismatch between box label, and pouch label, device hub/luer.The nurse cannot recall opening the box, therefore cannot confirm if the seals were broken before opening the box prior to use.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACIFIC XTREME
Type of Device
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key6338601
MDR Text Key67649965
Report Number9612164-2017-00184
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K103464
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2018
Device Catalogue NumberPCF030150130
Device Lot Number209917257
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2017
Date Device Manufactured08/12/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age88 YR
-
-