• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number JOEY
Device Problems Excess Flow or Over-Infusion (1311); Calibration Problem (2890)
Patient Problem Overdose (1988)
Event Date 01/13/2017
Event Type  malfunction  
Event Description
It was reported that at approximately 0400, it was noted that there was not enough formula remaining to feed patient until 1200, when the next 24 hour volume would be delivered to the unit.It is suspected that the patient was receiving more volume per hour than the 22 ml/hr that the pump was programmed for.Pump was removed from service and a new pump was ordered and feed was restarted.It was reported that upon evaluation, the pump was found to be out of calibration and will be sent back to the manufacturer.Reportedly, there was no harm to patient.It was also reported that 4 days prior to this event, this pump was evaluated by clinical engineering at this facility and returned to service with no issues identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN LP
444 mcdonnell blvd.
hazelwood MO 63042
MDR Report Key6338811
MDR Text Key67671488
Report Number6338811
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberJOEY
Other Device ID Number27196
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/02/2017
Device Age2 YR
Event Location Hospital
Date Report to Manufacturer02/02/2017
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age1 YR
Patient Weight5
-
-