Catalog Number HBD-W-12-13.5-15
Device Problem Hole In Material
Event Type  Malfunction  
Manufacturer Narrative

Concomitant products: cook dilation syringe, ds-60cc-s; boston scientific alliance 2 inflation device. Investigation evaluation: our laboratory evaluation of the product said to be involved confirmed the report. During a functional test, a cook ds-60cc-s syringe was filled with water and attached to the balloon inflation port. The syringe was placed onto an inflation handle, and negative pressure was applied to the balloon. After applying negative pressure, the balloon was inflated with water. When filling the balloon up with water, a leak was observed coming out of the proximal end of the balloon at the balloon joint. There was evidence of glue present at the balloon joint. The wire guide was included in the return and no part of the balloon was missing. A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis. The device history record for the lot number said to be involved was reviewed. A discrepancy or anomaly was not observed with the product that was released for distribution. Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting. Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis. This limits our ability to conclusively determine a cause. The additional information provided indicated that it is unknown if the user applied lubrication. A possible contributing factor to balloon material damage is failure to apply lubrication to the balloon with a lubricating agent. The instructions for use direct the user: ¿apply a lubricating agent to the balloon to facilitate passage through the endoscope accessory channel. " this activity will aid in endoscopic advancement and balloon preservation. The additional information provided indicated that it is unknown if the user provided negative pressure. Another possible contributing factor to balloon material damage is failure to apply negative pressure to the balloon dilator prior to advancement through the endoscope. The instructions for use direct the user: "to facilitate passage through the endoscope, apply negative pressure to the device. " the application of negative pressure will aspirate all residual air from the balloon and ease endoscopic advancement. Negative pressure will also aid in balloon preservation and optimize balloon performance. The instructions for use direct the user: ¿maintain balloon deflation with negative pressure and introduce into endoscope accessory channel, advancing in short increments until balloon is completely visualized endoscopically. ¿ damage to the balloon material can occur if the balloon is inflated prior to advancement through the endoscope or if the balloon is inflated while partially or fully inside the accessory channel of the endoscope. The instructions for use contain the following precaution: ¿do not pre-inflate the balloon. ¿ the instructions for use contain the following warning: ¿during dilation, do not inflate balloon beyond the maximum indicated inflation pressure. ¿ over inflation can cause damage to the balloon dilator. Another possible contributing factor is using a compromised inflation device to inflate the balloon. If the pressure reading of the inflation device is inaccurate, this could contribute to over inflation, possibly resulting in damage to the balloon material. Prior to distribution, all hercules 3 stage wire guided balloon esophageal-pyloric-colonic are subjected to a visual examination to ensure device integrity. A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment. Corrective action: a review of the complaint history was conducted. The likelihood of occurrence is considered rare. Corrective action is not warranted at this time based on the quality engineering risk assessment. Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.

 
Event Description

During an esophagogastroduodenoscopy (egd), the physician used a cook hercules 3 stage wireguided balloon esophageal-pyloric-colonic. There was a pinhole in the balloon.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHERCULES 3 STAGE WIREGUIDED BALLOON ESOPHAGEAL-PYLORIC-COLONIC
Type of DeviceKNQ, DILATOR, ESOPHAGEAL
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem , NC 27105
3367440157
MDR Report Key6345212
Report Number1037905-2017-00069
Device Sequence Number1
Product CodeKNQ
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,HEALTH
Reporter Occupation
Type of Report Initial
Report Date 02/21/2017
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received02/21/2017
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device Catalogue NumberHBD-W-12-13.5-15
Was Device Available For Evaluation? Yes
Is The Reporter A Health Professional? No
Was the Report Sent to FDA? No
Event Location Hospital
Date Manufacturer Received01/27/2017
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured12/01/2016
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 02/21/2017 Patient Sequence Number: 1
Treatment
OLYMPUS ENDOSCOPE, UNKNOWN MODEL
-
-