Model Number 809810 |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/30/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The reported issue was verified by the fsr on site.The device failed electrical safety by the reverse ungrounded leakage went above 100 a to 189 a.The fsr troubleshot to the heater on the a channel and confirmed it was the source of the problem.The heater was replaced and reverse ungrounded leakage was 69 a.Unit was operating within manufacture specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|
Event Description
|
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, heater cooler (hx2) failed electrical safety by the reverse ungrounded leakage going above 100 a to 189 a.There was no patient involvement.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician (pst) the resistance was measured between the heater's ground connection and heating elements.Both the main heating element and the secondary heating element measured approximately 2 megohms, 30 megohms was expected.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|