• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT Back to Search Results
Model Number 809810
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The reported issue was verified by the fsr on site.The device failed electrical safety by the reverse ungrounded leakage went above 100 a to 189 a.The fsr troubleshot to the heater on the a channel and confirmed it was the source of the problem.The heater was replaced and reverse ungrounded leakage was 69 a.Unit was operating within manufacture specifications.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The field service representative (fsr) reported that during preventive maintenance (pm) of the device, heater cooler (hx2) failed electrical safety by the reverse ungrounded leakage going above 100 a to 189 a.There was no patient involvement.
 
Manufacturer Narrative
During laboratory analysis, the product surveillance technician (pst) the resistance was measured between the heater's ground connection and heating elements.Both the main heating element and the secondary heating element measured approximately 2 megohms, 30 megohms was expected.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO HX2 TEMPERATURE MANAGEMENT SYSTEM
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS-HX2 TEMPERATURE MANAGEMENT
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
eileen dorsey
6200 jackson road
ann arbor, MI 48103
7347416074
MDR Report Key6349382
MDR Text Key68279738
Report Number1828100-2017-00090
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number809810
Device Catalogue Number809810
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/26/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-