MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) STEALTHSTATION S7 SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number S7
Device Problem Unstable (1667)
Patient Problem No Patient Involvement (2645)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Onsite investigation was preformed and the field representative suspected that the reported event was caused by the surgeon monitor and the external cable.Replacement of the aforementioned parts resolved the issue.No further issues were reported.Analysis of the returned external monitor cable could not confirm any failure as it passed bench testing and operated without any issues.The replaced monitor was not returned to manufacturer for analysis, therefore, no findings are possible.A full system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A site representative reported that the port, where the navigation system's surgeon monitor cable plugs into, felt loose and they sometimes lost video signal if that connection was manipulated.There was no patient present when this issue was identified.
 
Manufacturer Narrative
A medtronic representative reported that the monitor cord was replaced and not the monitor.The monitor was indicated as previously being replaced when it has not been replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION S7 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
peter verhey
826 coal creek circle
louisville, CO 80027
MDR Report Key6352611
MDR Text Key68127145
Report Number1723170-2017-00847
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994450944
UDI-Public00613994450944
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/23/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberS7
Device Catalogue Number9733856
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/26/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-