Model Number EX1150
Device Problems Power Conditioning Problem (1474); Noise, Audible (3273)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/26/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the transmitter lost power and then a loud pop sound was heard.The unit would be returned and replaced.
 
Manufacturer Narrative
Should have been (b)(6) 2017 rather than (b)(6) 2016 should have been "this report is to advise of an event that was observed during analysis".Should have been "reliability laboratory technician" rather than health professional.Other occupation should be "reliability laboratory technician" rather than na.Should have been "st.Jude medical crmd reliability lab" rather than consumer.Failure event observed during analysis.Analysis confirmed that the transmitter would not power up, the chip was damaged and the circuit board exhibited burn marks.
 
Event Description
The report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MERLIN@HOME, RF TELEMETRY BASIC
Type of Device
PACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6358502
MDR Text Key68510879
Report Number2017865-2017-01208
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 08/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/19/2016
Is the Reporter a Health Professional? No
Date Manufacturer Received08/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/05/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age75 YR
-
-