• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 7 SAGITTAL SAW INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 7 SAGITTAL SAW INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Catalog Number 7208000000
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem Abrasion (1689)
Event Date 02/02/2017
Event Type  Injury  
Manufacturer Narrative
Supplemental submitted to document device evaluation results.
 
Event Description
The customer reported that during a surgical procedure there was excessive oscillating of the saw that led to a rattling loose of the battery cover. This caused minor damage to the patellar tendon; however, the patellar tendon was repaired and the customer reported that lasting consequences to the patient as a result of this event are not likely.
 
Event Description
The customer reported that during a surgical procedure there was excessive oscillating of the saw that led to a rattling loose of the battery cover. This caused minor damage to the patellar tendon; however, the patellar tendon was repaired and the customer reported that lasting consequences to the patient as a result of this event are not likely.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYSTEM 7 SAGITTAL SAW
Type of DeviceINSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer (Section G)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
Manufacturer Contact
zachary baker
4100 east milham avenue
kalamazoo, MI 49001
2693237700
MDR Report Key6358598
MDR Text Key113532968
Report Number0001811755-2017-00672
Device Sequence Number1
Product Code HWE
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/23/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number7208000000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/23/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received08/29/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 02/24/2017 Patient Sequence Number: 1
-
-