Model Number 174001
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/17/2017
Event Type  malfunction  
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter: occurred during a laparoscopic rectal resection procedure.The device broke inside the abdomen but was removed in full.The part that broke was between the shaft and the jaw.No parts came loose from the device.To resolve the issue and complete the case, used another device.There was no patient harm.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The event report alleges the product was used in a surgical procedure.The plastic tip was cracked through the rivet point resulting in the graspers detaching.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Analysis concluded there were no assembly component related failures.Replication of the reported event can occur during rough handling of the device.The root cause of the observed damage was misuse of the product which would have caused or contributed to the reported incident.Should new information become available, the file will be re-opened and the investigation summary will be amended as appropriate.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO BABCOCK
Type of Device
MOTOR, SURGICAL INSTRUMENT, PNEUMATIC POWERED
Manufacturer (Section D)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer (Section G)
US SURGICAL PUERTO RICO
201 sabanetas industrial park
ponce PR 00716 4401
Manufacturer Contact
sharon murphy
60 middletown avenue
north haven, CT 06473
2034925267
MDR Report Key6366441
MDR Text Key68596236
Report Number2647580-2017-05022
Device Sequence Number1
Product Code GET
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K904578
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model Number174001
Device Catalogue Number174001
Device Lot NumberP6A0393X
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-