COOK INC NGAGE NITINOL STONE EXTRACTOR; FFL DISLODGER, STONE, BASKET, URETERAL, METAL Back to Search Results
Model Number N/A
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The event is currently under investigation.
 
Event Description
It was reported by the user facility that a patient was scheduled for an ureterorenoscopy procedure when the staff realized that the extracting element would not work as intended.The ngage nitinol stone extractor grasper was opened and prior to insertion checked for proper function.It was at that time the staff noticed that the gasper would not open.The device was placed back in the original packaging and another device was used to complete the procedure.No further information was provided.
 
Manufacturer Narrative
The device was returned the handle in the open position and the basket formation was partially open.A visual examination noted the support sheath is bent at the nose of the mlla.No kinks were noted in the basket sheath.A functional test was performed.The udh handle fully actuate the basket formation to open and closed positions.A review of the dhr showed 5 non-conformances associated with the complaint device lot number.Since some of the non-conformances are related to the complaint issue appropriate product and quality managers have been alerted.We will continue to monitor for similar complaints.Based on the provided information a definitive root cause cannot be established or reported at this time.We will notify the appropriate personnel and continue to monitor for similar complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NGAGE NITINOL STONE EXTRACTOR
Type of Device
FFL DISLODGER, STONE, BASKET, URETERAL, METAL
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key6367630
MDR Text Key68602361
Report Number1820334-2017-00407
Device Sequence Number1
Product Code FFL
UDI-Device Identifier00827002482975
UDI-Public(01)00827002482975(17)191128(10)7471557
Combination Product (y/n)N
Reporter Country CodeGB
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberNGE-022115
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-