MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Air Leak (1008)
Patient Problems Air Embolism (1697); ST Segment Elevation (2059)
Event Date 02/08/2017
Event Type  Injury  
Event Description
It was reported that during a cryoablation procedure after the sheath was advanced st elevation was observed as the sheath was advanced in the atrial septum. Spasm was suspected but a coronary angiography (cag) using a cag catheter showed there was no stenosis. It was believed that the st elevation was caused by air ingress; air in the ventricular apex and auricle was observed through fluoroscopy. Air aspiration was performed with a mapping catheter and the st elevation resolved. The procedure was then completed, and post-procedure, it was confirmed that the st elevation had resolved. No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: clinical data files were received and reviewed. The files cannot confirm the reported issue of ¿st elevation. ¿ the data files show the console controlling the pressure and flow properly without issues on the date of the event; at least 15 applications were completed with no issue. The sheath was not returned for investigation. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Event summary: the files cannot confirm the reported issue of ¿st elevation¿ and ¿air embolism¿. A good faith effort will be made to obtain the applicable information relevant to the report. If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameFLEXCATH ADVANCE STEERABLE SHEATH
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6373055
Report Number3002648230-2017-00099
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup,Followup,Followup
Report Date 02/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date10/13/2018
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number31099
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/13/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Outcome(s) Life Threatening;
-
-