• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012450-15
Device Problems Detachment Of Device Component (1104); Kinked (1339)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device was received.Investigation is not yet complete.A follow up report will be submitted with all relevant information.
 
Event Description
It was reported that during backloading onto the guide wire the 3.25 x 15 mm nc trek proximal balloon marker was noted to be bent.The device was not used; there was no patient interaction.A different nc trek balloon was used in the procedure without issue.There was no reported adverse patient effect.There was no reported clinically significant delay in the procedure.No additional information was provided.Return device analysis found that the inner member was separated 1mm proximal to the distal seal.The balloon and outer member were still intact.
 
Manufacturer Narrative
(b)(4).Evaluation summary: the device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda on march 17, 2017 [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6373444
MDR Text Key69165511
Report Number2024168-2017-01717
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08717648151934
UDI-Public(01)08717648151934(17)190731(10)60810G1
Combination Product (y/n)N
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number1012450-15
Device Lot Number60810G1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/27/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age59 YR
-
-