It was reported that the patient was referred for generator replacement.Device diagnostics at the time were within normal limits.Prior to the intended replacement surgery, high impedance was registered on the patient¿s vns.The patient was reportedly not aware of a potential device issue as he had not experienced any associated adverse events.The surgery was not performed at that time.Surgical intervention to address the high impedance has not occurred to date, and no additional pertinent information has been received to date.
|
The patient had surgery due to the high impedance.The lead pin was re-inserted into the generator, but the high impedance did not resolve.The physician then replaced both the generator and lead, which resolved the high impedance.The explanted products were not available from the hospital for return; therefore, no analysis could be performed.
|