Correction: a medtronic representative reported that some of these inaccuracies have resulted in patients coming back for re-do craniotomy.They were not able to provide how inaccurate the system was.The case was not delayed.They found at a later date that the tumor was not biopsied.They got results for normal brain tissue back from pathology.No patient demographics were provided despite multiple attempts from the surgeon and site staff.The following 3 related mdrs were submitted based on the above allegations of multiple patient events: 1723170-2017-01004, 1723170-2017-01001, 1723170-2017-01002.Additional information: a medtronic representative reported that he tested instruments and one probe was found to have slightly higher geometry error and was replaced (this probe was returned for evaluation and found to be fully functional and unrelated to the cause of the reported event).The rep found that after replacing chipped and scratched spheres, he was able to improve the instruments accuracy.He also found that "the locking mechanism on the mayfield (non-medtronic device) had play in it allowing the head to move" when he bumped in to the demo head after he had accuracy confirmed with frame, probe and demo head.Demo head moved about 1/4" caused about 2-3 mm inaccuracy.The reps checked all mayfields at site, found every one of them had slight play when it was locked.One of them had about double the movement than the others.It was recommended that the mayfields rep check out their devices.The rep also completed in-service training with the navigation coordinator at the site.The instructions for use (ifu) that accompanies the device contains warnings regarding the use of damaged spheres.
|