MAUDE Adverse Event Report: SYNTHES SELZACH 8.5MM MEDULLARY REAMER HEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES SELZACH 8.5MM MEDULLARY REAMER HEAD Back to Search Results
Catalog Number 352.085
Device Problems Break (1069); Material Fragmentation (1261)
Patient Problem Device Embedded In Tissue or Plaque (3165)
Event Date 02/03/2017
Event Type  Injury  
Manufacturer Narrative
Device was used for treatment, not diagnosis.Additional narrative: patient identifier is not available for reporting.Device is an instrument and is not implanted/explanted.The subject device is expected to be returned to the synthes manufacturer for evaluation but has not yet been received.(b)(6).The investigation could not be completed; no conclusion could be drawn, as no product was received.A device history record review was performed for the subject device lot.Manufacturing location: synthes (b)(4).Date of manufacture: may 26, 2014.The review showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.No non-conformances were generated during the production of the subject device.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes europe reported an event in (b)(6) as follows: it was reported that during the manual widening of the left femoral medullary cavity using an 8.5mm medullary reamer head, the cap of the proximal third ruptured in the medullary cavity, leaving three small metal fragments.There was a 45 minute surgical delay due to the reported event.The patient¿s postoperative status was reported as excellent.Concomitant parts reported: 1x flexible shaft, part unk, lot unk.This report is 1 of 1 for (b)(4).
 
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
8.5MM MEDULLARY REAMER HEAD
Type of Device
REAMER
Manufacturer (Section D)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ  CH2545
Manufacturer (Section G)
SYNTHES SELZACH
bohnackerweg 5
selzach CH254 5
SZ   CH2545
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6378532
MDR Text Key69091963
Report Number3000270450-2017-10073
Device Sequence Number1
Product Code HTO
UDI-Device Identifier07611819243699
UDI-Public(01)07611819243699(10)30106
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number352.085
Device Lot Number30106
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/26/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
UNKNOWN FLEXIBLE SHAFT, QTY 1
Patient Outcome(s) Required Intervention;
Patient Age17 YR
Patient Weight68
-
-