On using onmicut blade, metal shavings broken off into shoulder joint additional information received via email from the affiliate on 2-21-2017.
What type of procedure was this? arthroscopic subacromial decompression were all the shavings removed from the patient? all visible shavings were removed what was used to remove the shavings? suction did this delay the procedure, and if so how long? five to 10 minutes.
|
The complaint device was received and evaluated.
When observed under magnification, the inner shaft showed signs of friction between the two shafts.
The friction marks indicate excessive force, which could lead to metal shavings as reported, confirming the complaint.
A possible root cause would be improper use of blade/bur such as too much force, speed, blade/bur selection on tissue type.
A dhr review has been conducted and our results indicate that this batch of product was processed without incident related to this complaint and therefore there is no evidence of manufacturing anomalies on the paperwork reviewed.
Further, a review into the depuy synthes mitek complaints system revealed no similar complaint of any kind for this lot of devices that were released to distribution.
A white paper was performed previously on this failure; the result indicated that the amount of shedding for this product is acceptable.
At this point in time, no further action is warranted.
However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
Udi: (b)(4).
|