MEDTECH SA ROSA SURGICAL DEVICE; COMPUTER-ASSISTED SURGICAL DEVICE Back to Search Results
Model Number ROSA 2.5.8
Device Problem Loss of Data (2903)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 02/07/2017
Event Type  malfunction  
Event Description
During surgery planning phase, user was not able to load patient images on the device.User opened the images with the planning station and noticed that one was corrupted.This corrupted image was deleted and it was then possible for the user to load patient images on the device.
 
Manufacturer Narrative
Image set used on the rosa during the software failure was found to be unusable.Corrupted data was removed and image set was able to be loaded.The root cause is faulty image data from the hospital imaging system, unrelated to rosa.Rosa is not suspected as being deficient and did not contribute to data corruption.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROSA SURGICAL DEVICE
Type of Device
COMPUTER-ASSISTED SURGICAL DEVICE
Manufacturer (Section D)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR  34000
Manufacturer (Section G)
MEDTECH SA
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
Manufacturer Contact
elise lagacherie
zac eureka
900 rue du mas de verchant
montpellier, languedoc-roussillon 34000
FR   34000
67107740
MDR Report Key6383942
MDR Text Key69275990
Report Number3009185973-2017-00029
Device Sequence Number1
Product Code HAW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K101791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 07/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Medical Equipment Company Technician/Representative
Device Model NumberROSA 2.5.8
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-