Catalog Number 217863185 |
Device Problems
Material Discolored (1170); Scratched Material (3020)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/23/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
|
|
Event Description
|
Instrument is damaged.Update 3/7/2017 - per evaluation of returned device, reamer is rusting and has circular gouges.
|
|
Manufacturer Narrative
|
Examination of the submitted reamer confirmed the reported damage.The submitted reamer reveals evidence of significant circular scoring/wear and stains are visible near the etchings.A search of the complaint database did not reveal any trends for the 217863185 mbt rev cem stem reamer 13mm product code.The root cause is being attributed to extended service/wear out.Based on this investigation's determination of wear out as the root cause, no corrective action is being pursued.Continue to monitor per (b)(4).Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|