• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 MBT REV CEM STEM REAMER 13MM; KNEE INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 MBT REV CEM STEM REAMER 13MM; KNEE INSTRUMENT/TRIAL Back to Search Results
Catalog Number 217863185
Device Problems Material Discolored (1170); Scratched Material (3020)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/23/2017
Event Type  malfunction  
Manufacturer Narrative
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.
 
Event Description
Instrument is damaged.Update 3/7/2017 - per evaluation of returned device, reamer is rusting and has circular gouges.
 
Manufacturer Narrative
Examination of the submitted reamer confirmed the reported damage.The submitted reamer reveals evidence of significant circular scoring/wear and stains are visible near the etchings.A search of the complaint database did not reveal any trends for the 217863185 mbt rev cem stem reamer 13mm product code.The root cause is being attributed to extended service/wear out.Based on this investigation's determination of wear out as the root cause, no corrective action is being pursued.Continue to monitor per (b)(4).Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REV CEM STEM REAMER 13MM
Type of Device
KNEE INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6384062
MDR Text Key69271244
Report Number1818910-2017-14126
Device Sequence Number1
Product Code HTO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number217863185
Device Lot NumberA1204
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/03/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/15/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-