Model Number 302-20
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/16/2017
Event Type  malfunction  
Event Description
It was reported that a patient had a full system revision due to high impedance.The company representative who was present at the case confirmed the high impedance was identified from systems diagnostics on the day of surgery.The generator was replaced and high impedance was again found on the replacement generator after two diagnostics tests, and the lead was replaced as a result.After the lead was replaced diagnostics were reported to okay afterward.The explanted devices were discarded by the explant facility.Additional relevant information has not been received to-date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEAD MODEL 302
Type of Device
LEAD
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6399024
MDR Text Key69842791
Report Number1644487-2017-03365
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2010
Device Model Number302-20
Device Lot Number200580
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received02/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2007
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age66 YR
-
-