Model Number 104
Device Problem Detachment Of Device Component (1104)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/16/2017
Event Type  malfunction  
Event Description
It was reported that the septum plug for the negative lead pin fell out of a generator during replacement surgery.The physician was screwing in the positive lead pin when the negative pin's septum plug fell out.Reportedly, the septum plug fell out without having been touched.The surgeon was able to pop the septum plug back into place, and a backup generator was not available.The surgeon then decided to use the generator with the septum plug popped back into place and implanted it in the patient.No further relevant information has been received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 104
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS - HOUSTON
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS - HOUSTON
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6401681
MDR Text Key69856602
Report Number1644487-2017-03371
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 03/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/21/2018
Device Model Number104
Device Lot Number5134
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Date Manufacturer Received02/16/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/19/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age41 YR
-
-