MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-15
Device Problems Device Operational Issue (2914); Physical Property Issue (3008)
Patient Problem No Patient Involvement (2645)
Event Date 02/08/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of (b)(4).A (b)(4) field service representative was dispatched to the facility to investigate.The service representative tightened the guide sleeve (60-02-02) of the encoder to resolve the reported issue.Subsequent testing found no further issues and the unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.Evaluated on site by (b)(4) technician.
 
Event Description
(b)(4) received a report that the encoder of the centrifugal pump system with tubing clamp would not control the flow speed during priming.The user stated that the knob was found to be loose.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6404820
MDR Text Key69951929
Report Number9611109-2017-00175
Device Sequence Number1
Product Code DWA
UDI-Device Identifier04033817900986
UDI-Public010403381790098611150907
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 03/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-15
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/07/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-