Catalog Number UNKNOWN
Device Problems Complete Blockage (1094); Difficult to Remove (1528); Unintended Movement (3026); Insufficient Information (3190)
Patient Problems Anxiety (2328); No Information (3190)
Event Date 06/07/2010
Event Type  malfunction  
Manufacturer Narrative
Manufacturer ref# (b)(4).Catalog#: unknown but referred to as a cook günther tulip filter.(b)(4).Investigation is still in progress.
 
Event Description
Description according to short form complaint filed: it is alleged that "[pt] received a gunther tulip filter on (b)(6) 2010".Patient outcome: it is alleged that [pt] was injured without further explanation.Hospital and medical records have been requested but not yet provided.
 
Manufacturer Narrative
(b)(4).The event is currently under investigation.A supplemental report will be provided upon conclusion.
 
Event Description
This additional information was received on 04/06/2017 as follows: the patient allegedly received the device implant on (b)(6) 2010 due to pe with inability to anticoagulate following back surgery.An unsuccessful attempted removal allegedly occurred on (b)(6) 2010.The patient is alleging tilt, thrombus in filter, device is unable to be retrieved, and anxiety.
 
Manufacturer Narrative
Manufacturer ref # (b)(4).Exemption number e2016032.(b)(4).Corrected data based on new information received: adverse event to product problem, serious injury to malfunction.It has not been possible to further investigate or evaluate this alleged event based on the limited information provided to date via the operative note stating 'tilt, thrombus in filter, device unable to be retrieved, anxiety'.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Filter retrieval is occasionally difficult.This is well-known from published scientific literature where filter retrievals are referred to as simple vs.Complex.Several case reports published in scientific literature describe complex cases with successful endovascular filter retrievals using additional, advanced techniques.Filter tilt is a known risk in relation to filter implant reported in the published scientific literature and may occur during placement or during implanting period.Ivc thrombotic occlusion as an outcome for cook ivc filters is addressed in the published scientific literature.Ivc thrombotic occlusion is defined as the presence of an occluding thrombus in the ivc after filter insertion and documented by ultrasound (us), ct, mr imaging or venography; this may be symptomatic or asymptomatic.Unknown if the reported anxiety is directly related to the filter and unable to identify a corresponding failure mode at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer Contact
thomas hessner kirk
400 daniels way
bloomington, IN 47404
8004574500
MDR Report Key6412445
MDR Text Key70171853
Report Number3002808486-2017-00838
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090140
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup
Report Date 08/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Distributor Facility Aware Date03/13/2017
Date Manufacturer Received08/17/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient Age74 YR
Patient Weight73
-
-