The customer contacted the siemens customer care center (ccc) and allowed siemens remote service (srs) access.The customer stated that the instrument was the only dimension vista that the samples were run on.The customer performed patients' sample comparison between that dimension vista instrument and an alternate dimension vista instrument and the results were comparable.The customer ran quality controls (qc) for levels 1 and 3, resulting within range.The customer ran calibration, resulting acceptable.The customer stated that they believe that the original sample was mixed up with another sample.The cause of the discordant, falsely elevated tbil results is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
|
Discordant, falsely elevated total bilirubin (tbil) results were obtained on a neonatal patient sample on a dimension vista 500 instrument.The initial discordant result was reported to the maternity ward and the physician(s) questioned the result.A new sample from the infant drawn two hours later resulted lower.The result from the new sample was reported to the physician(s).The original sample was then repeated on the same dimension vista instrument, matching the initial discordant result.There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated tbil results.
|