The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110v (pump max).According to the technologist, the pump max and canister were incorrectly connected.However, during the procedure, blood backed up into the pump max.Therefore, use of the pump max was discontinued and the procedure was completed by other means.There was no report of an adverse effect to the patient.
|
Results: a pipe cleaner was inserted into the pump max vacuum port and blood was observed to be inside the penumbra system aspiration pump max 110 v (pump max).Conclusions: evaluation of the returned device revealed that there was blood inside the pump max.If the aspiration tubing is connected directly to the vacuum inlet rather than the canister supplied by penumbra, blood will likely enter the pump assembly.Penumbra pumps are visually and functionally inspected during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|