• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CYBERONICS, INC. PULSE GEN MODEL 103; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CYBERONICS, INC. PULSE GEN MODEL 103; GENERATOR Back to Search Results
Model Number 103
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Code Available (3191)
Event Date 02/28/2017
Event Type  Injury  
Event Description
It was reported that a patient had generator pocket revision surgery due to "pocket moved, lead pulling." the physician replaced the generator prophylactically.The physician believed that the generator migration and lead pulling was due to the patient shoveling snow and breaking the absorbable suture that was used to secure the generator.The surgery was done to prevent a lead fracture.The generator was received, but analysis has not been approved to date.
 
Event Description
The surgeon reported later that a non-absorbable suture was used to secure the generator, and non-absorbable sutures were used to close the incision at the skin level.Analysis on the generator was approved.The pulse generator diagnostics were as expected for the programmed parameters.A comprehensive automated electrical evaluation showed that the pulse generator performed according to functional specifications.There were no performance or any other type of adverse conditions found with the pulse generator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston, TX 77058
2812287200
MDR Report Key6434612
MDR Text Key70860276
Report Number1644487-2017-03468
Device Sequence Number1
Product Code LYJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/31/2015
Device Model Number103
Device Lot Number3747
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2017
Is the Reporter a Health Professional? Yes
Event Location Other
Date Manufacturer Received03/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age16 YR
-
-