• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. SARNS 8000 UNIVERSAL ROLLER PUMP; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER Back to Search Results
Model Number 16402
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 03/03/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the roller head was skipping.No other details regarding the nature of this event were provided.
 
Manufacturer Narrative
The field service representative (fsr) could not verify the pump to be skipping because the roller pump gave an overspeed 1 error when the motor was started.Troubleshooting found the pump driver printed circuit board assembly (pcba) to be the source of the problem.A failing pump driver pcba can cause pump to skip.The fsr also found that the fan that cools the pump driver pcba was not turning but would turn when touched.A failing cooling fan can cause the pump driver pcba to fail.The fsr replaced the pump driver pcba.The unit operated to the manufacturer's specifications.The suspect part will be returned to manufacturer for further testing.
 
Manufacturer Narrative
(b)(4).During laboratory analysis, the product surveillance technician (pst) inspected the circuit board and found overheating on the connector where the pump's motor was attached.The pst attached the board to a lab 8k pump test fixture and the poor connection to the motor caused the pump's motor to stop and start randomly.The pst connected the returned fan to a power supply and fan did not rotate when 12 volts was applied.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SARNS 8000 UNIVERSAL ROLLER PUMP
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ SARNS 8000 UNIVERSAL ROLLER
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
6200 jackson road
ann arbor MI 48103
Manufacturer Contact
katie hoyt
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key6438640
MDR Text Key71221041
Report Number1828100-2017-00148
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K953901
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16402
Device Catalogue Number16402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/13/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received06/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/07/1999
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-