The investigation has determined that a (b)(6) vitros anti hav igm result was obtained from a single patient sample tested on a vitros 3600 system.The definitive assignable cause could not be determined.No precision testing was performed to verify the performance of the vitros instrument at the time of the event and therefore an instrument issue cannot be entirely ruled out as contributing to the event.No historical quality control results were provided for the event so a vitros anti hav igm reagent performance issue also cannot be ruled out as contributing to the event.In addition, an unknown sample interferent, method to method difference or improper sample processing (sample centrifugation) cannot be ruled out as contributing factors.
|