BERCHTOLD GMBH & CO. KG CHROMOPHARE OR LIGHT CEILING-MOUNTED; SURGICAL LIGHT Back to Search Results
Catalog Number CH00000001
Device Problem Detachment Of Device Component (1104)
Patient Problem No Patient Involvement (2645)
Event Date 03/24/2017
Event Type  malfunction  
Manufacturer Narrative
It was reported that a d series light head allegedly detached from the spring arm.  a stryker field service technician (sfst) went onsite and found that the four screws that hold the light head to the cardanic were found to be backed out or broken.The sfst secured the three screws that backed out and replaced the one that was broken.Further investigation was performed and it was determined that the light has exceeded its expected lifetime.There was no patient involvement, injury or adverse consequence reported.
 
Event Description
It was reported that the light head allegedly detached from the spring arm.There was no patient involvement, injury or adverse consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHROMOPHARE OR LIGHT CEILING-MOUNTED
Type of Device
SURGICAL LIGHT
Manufacturer (Section D)
BERCHTOLD GMBH & CO. KG
ludwigstaler strasse 25
tuttlingen 78532
Manufacturer (Section G)
BERCHTOLD GMBH & CO. KG
ludwigstaler strasse 25
tuttlingen 78532
Manufacturer Contact
volker hornscheidt
ludwigstaler strasse 25
tuttlingen 78532
74611810
MDR Report Key6460432
MDR Text Key71728735
Report Number0008010153-2017-00020
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 04/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/05/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCH00000001
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/24/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-