Catalog Number C-HS-3045
Device Problems Difficult to Insert (1316); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4). A lot history record review was completed for the reported product lot number. There was no nonconformance recorded in the lot history. The device has not yet been returned to maquet cardiac surgery for evaluation. We are following up with the customer for the return of the device. A supplemental report will be submitted if the device is received.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system was not loaded into the delivery system properly. A replacement device was used to complete the procedure. The hospital did not report any patient effects.
 
Manufacturer Narrative
(b)(4). The device was returned to the factory for evaluation. Signs of clinical usage and no evidence of blood were observed. The delivery device was returned outside the loading device. The seal and tension spring assembly remained inside the loading device. The slide lock was dis-engaged and the plunger was not depressed on the delivery device. The seal was taken out from the loading device for inspection. The seal was observed to be intact with no cracks or delamination. The following measurements were taken; the inner delivery tube diameter was measured at. 198 in. The outer diameter was measured at. 224 in. The length of the delivery tube was measured at 2. 51 in. The values recorded were within the tolerance specifications. Based upon the received condition of the device, the reported complaint "failure to load" was confirmed.
 
Event Description
The hospital reported that during a coronary artery bypass procedure, hs iii proximal seal system was not loaded into the delivery system properly. A replacement device was used to complete the procedure. The hospital did not report any patient effects.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHS III PROXIMAL SEAL
Type of DeviceCLAMP, VASCULAR
Manufacturer (Section D)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer (Section G)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
Manufacturer Contact
45 barbour pond drive
wayne, NJ 07470
MDR Report Key6465603
MDR Text Key72229557
Report Number2242352-2017-00346
Device Sequence Number1
Product Code DXC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K080169
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 04/06/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date10/24/2017
Device Catalogue NumberC-HS-3045
Device Lot Number25129026
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/05/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received04/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-