MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION VISTA 1500; CLINICAL CHEMISTRY ANALYZER, Back to Search Results
Model Number DIMENSION VISTA 1500
Device Problem Mechanical Problem (1384)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2017
Event Type  malfunction  
Manufacturer Narrative
A siemens customer service engineer (cse) was dispatched to the customer site to inspect the analyzer.The cse found aliquot clots and probe detect errors during the aspiration.The cse noticed that the aliquot probe had gel and the instrument buffer tube static discharge plate was dirty, and cleaned it.The cse replaced the aliquot probe and auto-aligned it.The cse ran quick check, service methods on server 2 and system check, which passed.All results were within the acceptable standard deviations.The cause of the discordant, falsely elevated tbil result is unknown.The instrument is performing according to specifications.No further evaluation of the device is required.
 
Event Description
A discordant, falsely elevated total bilirubin (tbil) result was obtained on a patient sample on a dimension vista 1500 instrument.The discordant result was not reported to the physician(s).The sample was repeated on the same instrument, resulting lower.The sample was then tested on alternate dimension vista instrument, matching the repeated result from the original instrument.The repeated result was reported to the physician(s).There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated tbil result.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIMENSION VISTA 1500
Type of Device
CLINICAL CHEMISTRY ANALYZER,
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
500 gbc drive
po box 6101
newark DE 19714 6101
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS INC
registration #: 1226181
101 silvermine road
brookfield CT 06804
Manufacturer Contact
cassandra kocsis
511 benedict ave
tarrytown, NY 10591
9145242687
MDR Report Key6472798
MDR Text Key72340728
Report Number2517506-2017-00352
Device Sequence Number0
Product Code MQM
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051087
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2017
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDIMENSION VISTA 1500
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/16/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-