• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-751LNAB
Device Problem Failure to Deliver (2338)
Patient Problems Hypoglycemia (1912); Virus (2136); Vomiting (2144)
Event Date 04/01/2017
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
Customer reported via phone call to have been hospitalized on (b)(6) 2017 with blood glucose of 54 mg/dl.Customer had symptom of low blood glucose; customer was vomiting and had a stomach virus.Customer was hospitalized due to gastroparesis.Customer was still hospitalized at the time of call and was treated with injections.Customer was wearing insulin pump at the time of hospitalization.Customer also had high blood glucose of 300 mg/dl due to no delivery alarm.Troubleshooting could not be completed as insulin pump was removed for treatment and sets were thrown away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6473706
MDR Text Key72173926
Report Number3004209178-2017-40988
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial
Report Date 04/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/07/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-751LNAB
Device Catalogue NumberMMT-751LNAB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/04/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age27 YR
Patient Weight67
-
-