Dexcom was made aware on (b)(6) 2017, that on (b)(6) 2017, the patient experienced a missing sensor wire.The sensor was inserted in the patients arm.Urgent care physician stated that the patient was unsure if the wire was still in her skin.The patient went to urgent care on (b)(6) 2017 where x-rays were taken of the sensor location on the arm and the results were negative.The x-ray did not find any portion of the sensor wire in the patient's arm.The urgent care physician stated that the patient was not experiencing any discomfort or skin reactions.No additional patient or event information was provided.No product was provided for evaluation.The reported event could not be confirmed.A root cause cannot be determined.Labeling indicates: do not insert the sensor component of the dexcom system in a site other than the belly/abdomen (ages 2 years and older) or the upper buttocks (ages 2 to 17 years).The placement and insertion of the sensor component of the dexcom system is not approved for other sites.If placed in other areas, the dexcom system may not function properly.
|