Model Number U128
Device Problems Pacing Problem (1439); Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/09/2017
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete. This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that one day post implant it was noted the cardiac resynchronization therapy pacemaker (crt-p) was pacing in the left ventricle 99 percent, but pacing in the right ventricle 0 percent. It was found that the ventricular rate response was programmed off, once reprogrammed on there was 100 percent pacing in the right ventricle. No adverse patient effects were reported and the device remains in service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACCOLADE
Type of DeviceIMPLANTABLE PULSE GENERATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6479827
MDR Text Key72344530
Report Number2124215-2017-03051
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 02/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/14/2018
Device Model NumberU128
Other Device ID NumberVALITUDE X4
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Event Location No Information
Date Manufacturer Received02/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/11/2017 Patient Sequence Number: 1
Treatment
4674; 7741; U128
-
-