• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T177
Device Problems Over-Sensing (1438); Low impedance (2285); Ambient Noise Problem (2877)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that this implantable cardioverter defibrillator (icd) exhibited low out of range pacing impedances less than 200 ohms on the right ventricular (rv) channel.Episodes of noise oversensing were also observed, but the oversensing did not lead to pacing inhibition.A chest x-ray revealed no obvious rv lead damage.A revision procedure was performed.The rv lead was surgically abandoned and this icd was explanted and discarded.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6481402
MDR Text Key72409625
Report Number2124215-2017-04276
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
PMA/PMN Number
P960040/S039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/11/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/25/2009
Device Model NumberT177
Other Device ID NumberVITALITY 2 VR EL
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Manufacturer Received03/08/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/26/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0185; T177
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age70 YR
-
-