• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GUIDANT CRM CLONMEL IRELAND VITALITY 2; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number T167
Device Problem Premature Discharge of Battery (1057)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/01/2016
Event Type  Injury  
Manufacturer Narrative
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
Boston scientific received information that the patient called regarding this old implantable cardioverter defibrillator (icd) as it was replaced due to fast battery depletion.The icd was explanted.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Upon receipt at our post market quality assurance laboratory, additional information from product analysis indicates that the allegation was not confirmed.The battery status was end of life (eol).Elective replacement indicator (eri) was set before explant on (b)(6) 2016 and eol was set after explant.A review of the device memory noted no resets or fault codes.The device passed therapy verification and testing, thus no further analysis was performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALITY 2
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer (Section G)
GUIDANT CRM CLONMEL IRELAND
guidant ireland
clonmel, tipperary ireland
Manufacturer Contact
tim degroot
4100 hamline ave. n
st. paul, MN 
6515824786
MDR Report Key6485118
MDR Text Key72519787
Report Number2124215-2017-01893
Device Sequence Number1
Product Code KRG
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 02/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/14/2007
Device Model NumberT167
Other Device ID NumberVITALITY 2 DR EL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/26/2016
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received02/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/19/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
0184; 4472; T167
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age68 YR
-
-