AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012448-15
Device Problems Kinked (1339); Nonstandard Device (1420); Difficult to Remove (1528)
Patient Problem No Patient Involvement (2645)
Event Date 02/27/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.[medwatch # 2024168-2017-02310].
 
Event Description
It was reported that the protective sheath of the 2.75 x 15 mm nc trek was difficult to remove.Force was applied to remove the sheath and the dilatation catheter shaft became kinked.The device was not used and a second balloon dilatation catheter was used without issue.There was no patient involvement and no clinically significant delay in procedure reported.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda.[medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6486761
MDR Text Key72784484
Report Number2024168-2017-03072
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 05/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2019
Device Catalogue Number1012448-15
Device Lot Number60202G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/17/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2037-2039-2017
Patient Sequence Number1
-
-