Reportedly, while attempting to prepare a 6x200mm armada 18 percutaneous transluminal angioplasty (pta) catheter a negative was pulled; however, it was noticed that the hub was leaking (bubbles were noticed).The indeflator was disconnected from the device and the device was inspected.It was noticed that the hub was not formed correctly/broken; which didn't allow the devices to connect properly.The device was not used and the procedure was successfully completed with a new 6x200mm armada 18 pta.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.Although, initial testing of the returned device indicated that the balloon fully inflated to nominal pressure of 8 atmospheres without any leaks observed, additional testing was done and a leak was noted coming from the distal end of the hub at 6 atmospheres.
|