CYBERONICS, INC. PULSE GEN MODEL 102 GENERATOR Back to Search Results
Model Number 102
Device Problem Adverse Event Without Identified Device or Use Problem
Event Date 03/14/2017
Event Type  Injury  
Manufacturer Narrative

 
Event Description

It was reported by the patient that she felt like she was having a heart attack and questioned if the vns would make you feel like that. It was explained to the patient that vns does not make you feel like that. The patient took some of her husband's nitroglycerin, which made her feel better. She explained that she still felt that she had a knife stabbing her in the shoulder and her jaw hurt when using the vns magnet. The patient was urged multiple times during the call to speak with emergency care if she suspects a heart attack. The patient insisted on seeing a neurologist. It was later reported the vns was programmed off. The device history records for both the lead and the generator were reviewed and they had passed qc review prior to distribution. Attempts for further relevant information have been unsuccessful to date.

 
Event Description

Additional information was received from the physician's office. It was explained the patient had come in the month prior complaining of headache, left arm numbness, nausea, and chest pain. The device was programmed off and x-rays were performed. Nothing remarkable on the x-rays; the x-rays were normal. It was noted the patient came back a few days later and the physician stated since the patient was doing fine with the vns programmed off, they decided to leave it programmed off. No diagnostics were performed and it was not noted if the battery was low or not. A battery life calculation was performed showing the vns should have approximately 2 years remaining until neos = yes (near end of service), so it is not suspected the issues are due to a depleted battery. Additionally, the information available in the in-house programming history database was reviewed, and there were no anomalies observed with the patient's vns. However, the last diagnostic tests available were performed back in 2012.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePULSE GEN MODEL 102
Type of DeviceGENERATOR
Manufacturer (Section D)
CYBERONICS, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
CYBERONICS, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
njemile crawley
100 cyberonics blvd
suite 600
houston , TX 77058
2812287200
MDR Report Key6489520
Report Number1644487-2017-03608
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type COMPANY REPRESENTATIVE,CONSUM
Reporter Occupation PATIENT
Type of Report Initial,Followup
Report Date 04/13/2017
1 Device Was Involved in the Event
1 Patient Was Involved in the Event
Date FDA Received04/13/2017
Is This An Adverse Event Report? Yes
Is This A Product Problem Report? No
Device Operator LAY USER/PATIENT
Device EXPIRATION Date07/31/2010
Device MODEL Number102
Device LOT Number2214
Was Device Available For Evaluation? No
Is The Reporter A Health Professional? No
Was the Report Sent to FDA?
Event Location Other
Date Manufacturer Received04/13/2017
Was Device Evaluated By Manufacturer? Device Not Returned To Manufacturer
Date Device Manufactured10/22/2008
Is The Device Single Use? Yes
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient TREATMENT DATA
Date Received: 04/13/2017 Patient Sequence Number: 1
-
-