• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMEDA, INC. PURELY YOURS ULTRA; ELECTRIC BREAST PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMEDA, INC. PURELY YOURS ULTRA; ELECTRIC BREAST PUMP Back to Search Results
Model Number 24501879
Device Problem Fluid/Blood Leak (1250)
Patient Problem Burn(s) (1757)
Event Date 03/21/2017
Event Type  Injury  
Manufacturer Narrative
The returned product was tested per ameda engineering protocol to determine whether the allegation of leaking fluid could be confirmed.The returned product was assessed for indications of malfunction or thermal event.The returned product was assessed for functionality and met functional specifications.The battery fluid crust was observed inside the battery compartment.The returned product was assessed for functionality and met functional specifications.Additional testing supports that the misplacement of a battery can cause battery leakage.
 
Event Description
Customer contacted ameda, inc.On (b)(6) 2017 to report an incident that occurred that morning while pumping with her purely yours ultra breast pump while in the car.She states that she was pumping using battery power with 6 aa batteries installed in the pump base.She reports hearing a loud pop inside the pump base then noted dark fluid leaking from the battery compartment.Customer reports coming in contact with the fluid with her right hand and 3 fingers were burned.She reports her fingers were tingling, irritated and red.Customer washed with cool water and tingling resolved.She states the finger irritation and redness resolved 30 minutes after the incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PURELY YOURS ULTRA
Type of Device
ELECTRIC BREAST PUMP
Manufacturer (Section D)
AMEDA, INC.
485 half day rd.
suite 320
buffalo grove IL 60089
Manufacturer Contact
linda zager
485 half day rd.
suite 320
buffalo grove, IL 60089
8479642620
MDR Report Key6490160
MDR Text Key72739483
Report Number3009974348-2017-00260
Device Sequence Number1
Product Code HGX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K973501
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Remedial Action Replace
Type of Report Initial
Report Date 03/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number24501879
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/21/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/23/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age20 YR
-
-