ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP RF CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3262
Device Problems Telemetry Discrepancy (1629); Communication or Transmission Problem (2896)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/14/2017
Event Type  malfunction  
Event Description
It was reported that the asymptomatic patient presented to a right ventricular lead revision due to noise, low impedance and increased capture thresholds. The lead was capped and replaced. During the procedure, it was noted that the pulse generator was restricted to wand only telemetry. The device could be interrogated normally but telemetry was slow. The physician elected to explant and replace the pulse generator during the procedure. The patient was in stable condition during and post surgery.
 
Manufacturer Narrative
Final analysis found normal device characteristics.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUADRA ALLURE MP RF CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6491079
MDR Text Key72793645
Report Number2017865-2017-02289
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/06/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/30/2018
Device Model NumberPM3262
Device Lot NumberA000030181
Other Device ID Number05414734509091
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-