Model Number I7
Device Problems No Display/Image (1183); Device Stops Intermittently (1599); Device Displays Incorrect Message (2591)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/15/2017
Event Type  malfunction  
Manufacturer Narrative
On 03/16/2017 a medtronic representative performed a navigation system check-out, all areas passed. System performed as intended. On 04/11/2017 software engineering analysis completed. Patient logs were reviewed, however, provided no additional insight regarding the reported issue. No parts have been received by manufacturer for analysis.
 
Event Description
A medtronic representative reported that, while in a cranial procedure, the site's navigation system monitor touchscreen went black. Displayed was a "no input detected" message for approximately 10 seconds, then the monitor resumed normal function. The issue was identified pre-op. No further details regarding this issue were provided. The surgeon opted to continue and completed the procedure with the use of the navigation system. Delay in therapy was less than one hour. There was no impact on patient outcome.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSTEALTHSTATION I7 INTEGRATED NAVIGATION SYSTEM
Type of DeviceNEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
judith ericson
826 coal creek circle
louisville, CO 80027-9710
7208902187
MDR Report Key6491930
MDR Text Key72791143
Report Number1723170-2017-01038
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994653482
UDI-Public00613994653482
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050438
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial
Report Date 04/13/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberI7
Device Catalogue Number9734060
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/15/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-