LIVANOVA DEUTSCHLAND CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 60-02-50
Device Problem Device Operational Issue (2914)
Patient Problem No Patient Involvement (2645)
Event Date 03/28/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4) manufactures the centrifugal pump system with tubing clamp.The incident occurred in (b)(6).This medwatch report is being filed on behalf of (b)(4).A (b)(4) field service representative was dispatched to the facility to investigate.The service representative confirmed the reported issue and traced the issue to a loose guide sleeve on the encoder.The guide sleeve was tightened to resolve the reported issue.The knob and guide sleeve were cleaned and subsequent testing found no further issues.The unit was returned to service.A review of the dhr did not identify any deviations or non-conformities relevant to the reported issue.(b)(4) has not been made aware of any further issues with this device following the repair.Evaluated on site by (b)(4) technician.
 
Event Description
(b)(4) received a report that the control knob of the centrifugal pump system with tubing clamp did not work correctly during priming.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIFUGAL PUMP SYSTEM WITH TUBING CLAMP
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich, 80939
GM   80939
Manufacturer Contact
joan ceasar
14401 w. 65th way
arvada, CO 80004
2812287260
MDR Report Key6492896
MDR Text Key72786731
Report Number9611109-2017-00317
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032213
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date 04/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number60-02-50
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-