Model Number PM3222
Device Problem Under-Sensing (1661)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/07/2017
Event Type  malfunction  
Event Description
It was reported that a merlin. Net transmission revealed the right atrial lead and pulse generator exhibited undersensing. The left ventricular lead exhibited loss of capture. The patient was brought into the clinic and an x-ray was performed which revealed that the left ventricular lead had dislodged. The physician elected to increase the device output. The patient was asymptomatic and would continue to be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameALLURE RF
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6494248
MDR Text Key72841655
Report Number2017865-2017-02313
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation
Type of Report Initial
Report Date 04/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date09/30/2016
Device Model NumberPM3222
Device Lot NumberA000003676
Other Device ID Number05414734507905
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/07/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/22/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/14/2017 Patient Sequence Number: 1
Treatment
1258T/86, BLG044319; 1699TC/46, EN028053
-
-