Model Number AU00T0
Device Problem Unintended Movement (3026)
Patient Problem Capsular Bag Tear (2639)
Event Date 04/04/2017
Event Type  Injury  
Manufacturer Narrative
Evaluation summary: the device was not returned for evaluation. Product history records were reviewed and documentation indicated the product met release criteria. Root cause has not been identified. There have been no other complaints reported in the lot number. Additional information has been requested. (b)(4).
 
Event Description
A healthcare professional reported a violent unfolding of an intraocular lens (iol) during implant while using a preloaded delivery system. The patient experienced a capsular rupture that resulted in a vitrectomy. Additional information has been requested but not received to date.
 
Manufacturer Narrative
Viscoelastic was not provided. It is unknown if a qualified product was used. Material properties of non-qualified ovds may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes. (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameACRYSOF IQ SINGLEPIECE IOL W/ULTRASERT DELIVERY SYSTEM
Type of DeviceLENS, GUIDE, INTRAOCULAR
Manufacturer (Section D)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LTD. - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
rita lopez
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514846
MDR Report Key6498065
MDR Text Key72940256
Report Number1119421-2017-00518
Device Sequence Number1
Product Code KYB
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P930014
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 05/08/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Expiration Date05/31/2019
Device Model NumberAU00T0
Device Catalogue NumberAU00T0.245
Device Lot Number12468170
Other Device ID Number00380652358361
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/02/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/16/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 04/17/2017 Patient Sequence Number: 1
-
-