SMITH & NEPHEW, INC. SVC REPL,MDU, HAND CNTRL, PWRMX SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 72200616S
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/1901
Event Type  malfunction  
Manufacturer Narrative
Device investigation narrative - one service replacement powermax elite motor drive unit, part number 72200616s was received on march 27, 2017 and confirmed to be serial number (b)(4). Complaint of overheating and motor stall error was confirmed. Cause of overheating and errors is a corroded motor/gearbox. The motor/gearbox was removed from the housing. The gearbox was removed from motor. The cable assembly and motor tested good. The gearbox was found to be jammed and corroded internally. A review of the manufacturing records shows that this unit was released to distribution on or about january 25, 2017. The root cause of this event was identified to be associated with corrosion of the motor and gearbox assembly. A motor stall condition will result in increased current draw from the control unit which will heat the motor and hand piece housing. Factors which can contribute to gearbox corrosion include cleaning and sterilization methods and the chemicals involved. A corrective action has been initiated to mitigate future recurrence of similar events. (b)(4).
 
Event Description
It was reported that the mdu was getting hot and then quit working. No patient or user impact was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSVC REPL,MDU, HAND CNTRL, PWRMX
Type of DeviceSAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107 6512
Manufacturer Contact
james gonzales
7000 west william cannon drive
austin, TX 78735
MDR Report Key6499827
MDR Text Key73033493
Report Number1643264-2017-00213
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation
Type of Report Initial
Report Date 04/18/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/18/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue Number72200616S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/27/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/21/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/24/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-